MIC calibration for system EC-912

MIC Smoke Measuring Equipment, type EC-012

General

According to requirements in common quality control systems, all test equipment for validating measurements shall be included in a calibration program

DELTA offers calibration and maintenance services for MIC Smoke Measuring Equipment, type EC-012.

A MIC calibration consists of:

Pre-cleaning measurement
Wipe test of radioactive source (leak prevention)
Cleaning of MIC chamber & Alpha source
Replacement of air filter in flow control unit
Calibration against the MIC reference chamber. MIC chambers with deviations higher than 3% from MIC World Reference Chamber is adjusted.

Results

The results of the calibration are presented in two documents: “Pre cleaning measurement” and “Certificate of Calibration”.

DELTA reference material complies with requirements of EN 54-7:2000/A2:2006 Annex M. Calibration is traceable to DFM (dimension) and NPLA (temperature, humidity and pressure).

MIC/Mirex calibrations are not a part of DELTA’s accredited calibration services.

Please note

For MIC calibration service you need to send us the following parts: EC-912-10 MIC measuring ionization chamber, EC-912-20 MIC control unit and EC-912-40 MIC Flow Control Unit. In order to minimise shipment costs no other parts are needed.

To ensure sufficient insurance and a smooth and regular handling in the customs, please use the price of 10,000 EUR for one MIC system on your proforma invoices.